CT Scanner
Completed

Toshiba Aquilion 64 TSX-101A 2025 Internal Report

Share

Diagnosis

The CT imaging system was unable to complete its operational readiness sequence, preventing diagnostic procedures. This issue was identified as originating from a malfunction within the equipment's critical thermal management system.

Work Performed

Initial service involved optimizing the thermal fluid within the cooling unit. Following a recurrence of the issue, a comprehensive diagnostic revealed a deeper problem with the internal circulation mechanism, leading to the replacement of the entire thermal management module. This fully restored the system to manufacturer specifications, ensuring reliable and precise operation.